Jump to content Jump to navigation menu

Single-channel microlitre pipette Acura® manual 810 for dilutions

From  Socorex
Icon_Autoklavierbar All autoclavable products in our catalogue are identified by this symbol. Please also note any additional instructions when autoclaving.
Autoclavable: yes.
Volume


Stem pipette for 1:10 dilutions

Air displacement pipette with two pre-calibrated fixed volumes in one piston stroke allows sequential delivery of 1 and 0.1 ml of the same liquid. For more flexibility, the volumes can be used either individually or in combination for dilutions.

The metal cone provides a perfect seat for long PP stem tips, ideal for taking samples from narrow vessels Stomacher®- bags and other blending bags. The Acura® manual 810 is an ideal alternative to conventional graduated pipettes for serial 1:10 dilutions in bacteriology.
Product details

€622.45/Pack Qty. 

excl. VAT. | 1 unit(s) per Pack Qty.

Art. No. HK00.1

In procurement
from 6 Pack Qty. €591.33/Pack Qty.
from 24 Pack Qty. €560.20/Pack Qty.

Product details

Autoclavable: yes.

Stem pipette for 1:10 dilutions

Air displacement pipette with two pre-calibrated fixed volumes in one piston stroke allows sequential delivery of 1 and 0.1 ml of the same liquid. For more flexibility, the volumes can be used either individually or in combination for dilutions.

The metal cone provides a perfect seat for long PP stem tips, ideal for taking samples from narrow vessels Stomacher®- bags and other blending bags. The Acura® manual 810 is an ideal alternative to conventional graduated pipettes for serial 1:10 dilutions in bacteriology.

Properties:
  • Two pre-calibrated fixed volumes – no adjustment necessary
  • The two volumes 1 and 0.1 ml can be calibrated separately
  • Easy performance of all functions
  • Excellent ergonomics, lightweight
  • Metal shaft allows the use of long stem tips
  • Justip™ system to adjust the height of the tip ejector
  • Simple, efficient ejection of stem tips
  • Replaceable PE protective filter

Technical specifications:

Volume: 100 μl:
  • Inaccuracy (E %): <±2,0
  • Imprecision (CV %): <±2,5

Volume: 1000 μl:
  • Inaccuracy (E %): <±0,5
  • Imprecision (CV %): <±0,4

The values were achieved with distilled water at a constant (±0.5 °C) ambient temperature between 20 and 25 °C, in accordance with. EN ISO 8655.

Single-channel microlitre pipette Acura® manual 810 for dilutions 

Not a medical device / Not an IVD product
Technical Information
Version Single-channel, fixed volume 
Volume 100 / 1000 µl
Single-channel microlitre pipette Acura® manual 810 for dilutions
Selected quantity:   0
  1. Subtotal:  0.00
Art. No. Volume Pack Qty. Price Quantity
HK00.1 100 / 1000 µl 1 unit(s)

€622.45

In stock
Available
In procurement
No longer available
Delivery date currently unknown
Selected quantity:   0
  1. Subtotal:  0.00

Downloads / MSDS