Compact system for the production of contamination-free water from mains water. The device is connected directly to the mains water supply.
Simultaneous provision of ultrapure and pure water qualities:
- ASTM Type I (resistance of 18.2 MΩ x cm)
- ASTM Type II (resistance of 10-15 MΩ x cm)
Easy to use, space-saving design and low follow-up costs. The pre-treatment cartridge is separate from the ultrapure water cartridge. Only the used cartridge need be replaced, when necessary.
Applications:
Analytical chemistry:
HPLC, GC, GC-MS, ICP-MS, AA, TOC measurements, IC
Molecular biology and microbiology:
Cell/tissue culture, PCR, DNA sequencing, electrophoresis
Routine tasks in the laboratory
Preparation and dilution of buffers and reagents, rinsing of laboratory glass instruments, supply of autoclaves, analysers and other laboratory instruments
- Combined cartridge (reverse osmosis membrane/activated charcoal) for pre-treatment of feed water
- Direct tapping of type I ultrapure water through a sterile end filter (0.2 µm)
- Integrated 6-litre tank for storing type II pure water
- Guaranteed pure water quality: the recirculation pump and sterile air filter protect from microbial contamination
- Additional pressure outlet on ultrapure water tank for supplying downstream devices
- Simple and economical change of expendables as treatment cartridges can be replaced separately
- Space-saving mounting on laboratory bench or on the wall
Drinking water requirements: | |
Drinking water to DIN 2000 | Manganese and iron content: <0.05 mg/l |
Chlorine concentration: <0.1 mg/l | |
Colloid index: <3 | |
Feed water temperature: 2-35 °C | |
Feed water pressure: 1-6 bar |
Pure and ultrapure water system Smart2Pure series
With UV lamp (wave length 254 and 185 nm), which inactivates bacteria, oxidises organic compounds and enables germ-free ultrapure water quality with a TOC concentration <5 ppb.Typical applications: Anorganic and organic trace analysis, HPLC, ICP-MS, IC, TOC analysis.
Type | Smart2Pure UV 3 l/h |
Conductivity of high-purity water, type I | 0.055 µS/cm |
Conductivity of pure water, type II | 0,067-0,1 µS/cm |
Resistance of high-purity water, type I | 18.2 MΩ*cm |
Resistance of pure water, type II | 10-15 MΩ*cm |
Withdrawal capacity of high-purity water, type I | 1 l/min |
Withdrawal capacity of pure water, type II | 3 l/h |
Bacteria content of high-purity water, type I | <1 KbE/ml |
TOC value of high-purity water, type 1 | 1-5 ppb |
Particle content of high-purity water, type I | <1 particle per ml (particle size <0.2 μm) |
Width | 305 mm |
Depth | 400 mm |
Height | 545 mm |
Weight | 22 kg |
Connection |
Feedwater: hose (length 1.5 m) with Ø outside 8 mm and 3/4″ Waste water: 1/4″ or Ø outside 8 mm, hose (length 2 m), pressure-free discharge |
Mains connection | 100-240 V, 50/60 Hz |
- Subtotal: 0.00
Order No. | Type | Pack Qty. | Price | Quantity | ||
---|---|---|---|---|---|---|
HE81.1 | Smart2Pure UV 3 l/h | 1 unit(s) |
€5,768.45 |
|
||
HE82.1 | Smart2Pure UV 6 l/h | 1 unit(s) |
€7,078.90 |
|
||
In stock
In procurement
Not available
Delivery date currently unknown
|
- Subtotal: 0.00
Downloads / MSDS
General information
Further attractive products to complete your chromatography laboratory can be found on our Chromatography page!